Τετάρτη 9 Αυγούστου 2017

Use of noise cancellation earphones in out-of-booth audiometric evaluations.

Related Articles

Use of noise cancellation earphones in out-of-booth audiometric evaluations.

Int J Audiol. 2017 Aug 08;:1-8

Authors: Clark JG, Brady M, Earl BR, Scheifele PM, Snyder L, Clark SD

Abstract
OBJECTIVE: To assess the utility of noise cancelation earphones (NCE) in audiometric evaluations.
DESIGN: Degree of noise reduction of Bose QuietComfort 15 NCE was assessed through probe-microphone measures and sound-field audiometry. Occlusion effects from NCE were assessed for potential effects on bone-conduction thresholds.
STUDY SAMPLE: Twenty participants were tested to determine average occlusion effect values during bone-conduction testing with and without NCE. Noise reduction values of the NCE were assessed on a single subject through probe-microphone measures and sound-field testing.
RESULTS: NCE sufficiently reduced ambient noise to levels acceptable for air-conduction testing as well as for bone-conduction testing for most patients when adding minimal adjustment to acceptable levels as outlined by the ANSI S3.1-1999 standard. In addition, NCE did not create a clinically significant change in the occlusion effect for bone-conduction testing.
CONCLUSION: NCE placed over insert earphones provide a sound pressure level at the tympanic membrane that is below ANSI standards for routine air-conduction testing and result in sufficient ambient noise reduction for bone-conduction testing with most patients. There is no clinically significant occlusion effect from NCE during routine bone-conduction audiometry. These findings support the utility of using NCE for offsite audiometric testing.

PMID: 28783994 [PubMed - as supplied by publisher]



from #Audiology via ola Kala on Inoreader http://ift.tt/2vOqoFq
via IFTTT

Use of noise cancellation earphones in out-of-booth audiometric evaluations.

Related Articles

Use of noise cancellation earphones in out-of-booth audiometric evaluations.

Int J Audiol. 2017 Aug 08;:1-8

Authors: Clark JG, Brady M, Earl BR, Scheifele PM, Snyder L, Clark SD

Abstract
OBJECTIVE: To assess the utility of noise cancelation earphones (NCE) in audiometric evaluations.
DESIGN: Degree of noise reduction of Bose QuietComfort 15 NCE was assessed through probe-microphone measures and sound-field audiometry. Occlusion effects from NCE were assessed for potential effects on bone-conduction thresholds.
STUDY SAMPLE: Twenty participants were tested to determine average occlusion effect values during bone-conduction testing with and without NCE. Noise reduction values of the NCE were assessed on a single subject through probe-microphone measures and sound-field testing.
RESULTS: NCE sufficiently reduced ambient noise to levels acceptable for air-conduction testing as well as for bone-conduction testing for most patients when adding minimal adjustment to acceptable levels as outlined by the ANSI S3.1-1999 standard. In addition, NCE did not create a clinically significant change in the occlusion effect for bone-conduction testing.
CONCLUSION: NCE placed over insert earphones provide a sound pressure level at the tympanic membrane that is below ANSI standards for routine air-conduction testing and result in sufficient ambient noise reduction for bone-conduction testing with most patients. There is no clinically significant occlusion effect from NCE during routine bone-conduction audiometry. These findings support the utility of using NCE for offsite audiometric testing.

PMID: 28783994 [PubMed - as supplied by publisher]



from #Audiology via ola Kala on Inoreader http://ift.tt/2vOqoFq
via IFTTT

Use of noise cancellation earphones in out-of-booth audiometric evaluations.

Related Articles

Use of noise cancellation earphones in out-of-booth audiometric evaluations.

Int J Audiol. 2017 Aug 08;:1-8

Authors: Clark JG, Brady M, Earl BR, Scheifele PM, Snyder L, Clark SD

Abstract
OBJECTIVE: To assess the utility of noise cancelation earphones (NCE) in audiometric evaluations.
DESIGN: Degree of noise reduction of Bose QuietComfort 15 NCE was assessed through probe-microphone measures and sound-field audiometry. Occlusion effects from NCE were assessed for potential effects on bone-conduction thresholds.
STUDY SAMPLE: Twenty participants were tested to determine average occlusion effect values during bone-conduction testing with and without NCE. Noise reduction values of the NCE were assessed on a single subject through probe-microphone measures and sound-field testing.
RESULTS: NCE sufficiently reduced ambient noise to levels acceptable for air-conduction testing as well as for bone-conduction testing for most patients when adding minimal adjustment to acceptable levels as outlined by the ANSI S3.1-1999 standard. In addition, NCE did not create a clinically significant change in the occlusion effect for bone-conduction testing.
CONCLUSION: NCE placed over insert earphones provide a sound pressure level at the tympanic membrane that is below ANSI standards for routine air-conduction testing and result in sufficient ambient noise reduction for bone-conduction testing with most patients. There is no clinically significant occlusion effect from NCE during routine bone-conduction audiometry. These findings support the utility of using NCE for offsite audiometric testing.

PMID: 28783994 [PubMed - as supplied by publisher]



from #Audiology via ola Kala on Inoreader http://ift.tt/2vOqoFq
via IFTTT

Use of noise cancellation earphones in out-of-booth audiometric evaluations.

Related Articles

Use of noise cancellation earphones in out-of-booth audiometric evaluations.

Int J Audiol. 2017 Aug 08;:1-8

Authors: Clark JG, Brady M, Earl BR, Scheifele PM, Snyder L, Clark SD

Abstract
OBJECTIVE: To assess the utility of noise cancelation earphones (NCE) in audiometric evaluations.
DESIGN: Degree of noise reduction of Bose QuietComfort 15 NCE was assessed through probe-microphone measures and sound-field audiometry. Occlusion effects from NCE were assessed for potential effects on bone-conduction thresholds.
STUDY SAMPLE: Twenty participants were tested to determine average occlusion effect values during bone-conduction testing with and without NCE. Noise reduction values of the NCE were assessed on a single subject through probe-microphone measures and sound-field testing.
RESULTS: NCE sufficiently reduced ambient noise to levels acceptable for air-conduction testing as well as for bone-conduction testing for most patients when adding minimal adjustment to acceptable levels as outlined by the ANSI S3.1-1999 standard. In addition, NCE did not create a clinically significant change in the occlusion effect for bone-conduction testing.
CONCLUSION: NCE placed over insert earphones provide a sound pressure level at the tympanic membrane that is below ANSI standards for routine air-conduction testing and result in sufficient ambient noise reduction for bone-conduction testing with most patients. There is no clinically significant occlusion effect from NCE during routine bone-conduction audiometry. These findings support the utility of using NCE for offsite audiometric testing.

PMID: 28783994 [PubMed - as supplied by publisher]



from #Audiology via ola Kala on Inoreader http://ift.tt/2vOqoFq
via IFTTT

RgIA4 Potently Blocks Mouse α9α10 nAChRs and Provides Long Lasting Protection against Oxaliplatin-Induced Cold Allodynia.

Related Articles

RgIA4 Potently Blocks Mouse α9α10 nAChRs and Provides Long Lasting Protection against Oxaliplatin-Induced Cold Allodynia.

Front Cell Neurosci. 2017;11:219

Authors: Christensen SB, Hone AJ, Roux I, Kniazeff J, Pin JP, Upert G, Servent D, Glowatzki E, McIntosh JM

Abstract
Transcripts for α9 and α10 nicotinic acetylcholine receptor (nAChR) subunits are found in diverse tissues. The function of α9α10 nAChRs is best known in mechanosensory cochlear hair cells, but elsewhere their roles are less well-understood. α9α10 nAChRs have been implicated as analgesic targets and α-conotoxins that block α9α10 nAChRs produce analgesia. However, some of these peptides show large potency differences between species. Additionally several studies have indicated that these conotoxins may also activate GABAB receptors (GABABRs). To further address these issues, we cloned the cDNAs of mouse α9 and α10 nAChR subunits. When heterologously expressed in Xenopus oocytes, the resulting α9α10 nAChRs had the expected pharmacology of being activated by acetylcholine and choline but not by nicotine. A conotoxin analog, RgIA4, potently, and selectively blocked mouse α9α10 nAChRs with low nanomolar affinity indicating that RgIA4 may be effectively used to study murine α9α10 nAChR function. Previous reports indicated that RgIA4 attenuates chemotherapy-induced cold allodynia. Here we demonstrate that RgIA4 analgesic effects following oxaliplatin treatment are sustained for 21 days after last RgIA4 administration indicating that RgIA4 may provide enduring protection against nerve damage. RgIA4 lacks activity at GABAB receptors; a bioluminescence resonance energy transfer assay was used to demonstrate that two other analgesic α-conotoxins, Vc1.1 and AuIB, also do not activate GABABRs expressed in HEK cells. Together these findings further support the targeting of α9α10 nAChRs in the treatment of pain.

PMID: 28785206 [PubMed]



from #Audiology via ola Kala on Inoreader http://ift.tt/2wGYuHQ
via IFTTT

RgIA4 Potently Blocks Mouse α9α10 nAChRs and Provides Long Lasting Protection against Oxaliplatin-Induced Cold Allodynia.

Related Articles

RgIA4 Potently Blocks Mouse α9α10 nAChRs and Provides Long Lasting Protection against Oxaliplatin-Induced Cold Allodynia.

Front Cell Neurosci. 2017;11:219

Authors: Christensen SB, Hone AJ, Roux I, Kniazeff J, Pin JP, Upert G, Servent D, Glowatzki E, McIntosh JM

Abstract
Transcripts for α9 and α10 nicotinic acetylcholine receptor (nAChR) subunits are found in diverse tissues. The function of α9α10 nAChRs is best known in mechanosensory cochlear hair cells, but elsewhere their roles are less well-understood. α9α10 nAChRs have been implicated as analgesic targets and α-conotoxins that block α9α10 nAChRs produce analgesia. However, some of these peptides show large potency differences between species. Additionally several studies have indicated that these conotoxins may also activate GABAB receptors (GABABRs). To further address these issues, we cloned the cDNAs of mouse α9 and α10 nAChR subunits. When heterologously expressed in Xenopus oocytes, the resulting α9α10 nAChRs had the expected pharmacology of being activated by acetylcholine and choline but not by nicotine. A conotoxin analog, RgIA4, potently, and selectively blocked mouse α9α10 nAChRs with low nanomolar affinity indicating that RgIA4 may be effectively used to study murine α9α10 nAChR function. Previous reports indicated that RgIA4 attenuates chemotherapy-induced cold allodynia. Here we demonstrate that RgIA4 analgesic effects following oxaliplatin treatment are sustained for 21 days after last RgIA4 administration indicating that RgIA4 may provide enduring protection against nerve damage. RgIA4 lacks activity at GABAB receptors; a bioluminescence resonance energy transfer assay was used to demonstrate that two other analgesic α-conotoxins, Vc1.1 and AuIB, also do not activate GABABRs expressed in HEK cells. Together these findings further support the targeting of α9α10 nAChRs in the treatment of pain.

PMID: 28785206 [PubMed]



from #Audiology via ola Kala on Inoreader http://ift.tt/2wGYuHQ
via IFTTT

Τρίτη 8 Αυγούστου 2017

Auditory Training for Adults Who Have Hearing Loss: A Comparison of Spaced Versus Massed Practice Schedules

Purpose
The spacing effect in human memory research refers to situations in which people learn items better when they study items in spaced intervals rather than massed intervals. This investigation was conducted to compare the efficacy of meaning-oriented auditory training when administered with a spaced versus massed practice schedule.
Method
Forty-seven adult hearing aid users received 16 hr of auditory training. Participants in a spaced group (mean age = 64.6 years, SD = 14.7) trained twice per week, and participants in a massed group (mean age = 69.6 years, SD = 17.5) trained for 5 consecutive days each week. Participants completed speech perception tests before training, immediately following training, and then 3 months later. In line with transfer appropriate processing theory, tests assessed both trained tasks and an untrained task.
Results
Auditory training improved the speech recognition performance of participants in both groups. Benefits were maintained for 3 months. No effect of practice schedule was found on overall benefits achieved, on retention of benefits, nor on generalizability of benefits to nontrained tasks.
Conclusion
The lack of spacing effect in otherwise effective auditory training suggests that perceptual learning may be subject to different influences than are other types of learning, such as vocabulary learning. Hence, clinicians might have latitude in recommending training schedules to accommodate patients' schedules.

from #Audiology via xlomafota13 on Inoreader http://article/doi/10.1044/2017_JSLHR-H-16-0154/2648749/Auditory-Training-for-Adults-Who-Have-Hearing-Loss
via IFTTT