Purpose
The purpose of this report is to describe the equipment setup and techniques for successfully testing behavioral thresholds in young children with cochlear implants (CIs) using telepractice. We will also discuss challenges associated with pediatric CI programming that are unique to the use of distance technology, and we will describe ways to overcome those challenges. Last, we will review the results from 2 recent studies specifically targeted toward testing behavioral thresholds in young children with CIs.
Method
Conditioned play audiometry or visual reinforcement audiometry was used to measure behavioral thresholds (T levels) for 35 young children with CIs (n = 19 for conditioned play audiometry and n = 16 for visual reinforcement audiometry). Participants were tested in the traditional in-person condition and in the remote condition using an AB-BA study design over 2 visits.
Results
There was no significant difference in T levels between the in-person and remote conditions, indicating that it is feasible to test young children using conventional pediatric testing procedures via remote technology. The primary challenges encountered were in regard to proper camera and video monitor placement at the remote site and the timing of communication between the audiologist and test assistant.
Conclusions
The results from studies to date suggest that distance technology can be used successfully to program CI sound processors for young children using standard, age-appropriate testing techniques. The alternative of remote testing has substantial implications for reducing time and travel burdens for families, potentially leading to the construction of appropriate maps for young children with CIs in a timelier manner.from #Audiology via ola Kala on Inoreader https://ift.tt/2Q9oFV1
via IFTTT
Δεν υπάρχουν σχόλια:
Δημοσίευση σχολίου