Comparison Study of Percutaneous Osseointegrated Bone Conduction Device Complications When Using the 9 mm Abutment Versus 6 mm Abutment at Initial Implantation

Objectives: To assess differences in the incidence, type, and management of complications encountered with implantation of percutaneous osseointegrated bone conduction devices when using a 9 mm abutment versus 6 mm abutment at initial implantation. Study Design: Retrospective cohort study. Methods: One hundred thirty consecutive patients between January 2010 and December 2011 underwent single-stage percutaneous osseointegrated bone conduction device implantation using a 9 or 6 mm abutment. Clinical outcomes assessed for the two groups included the incidence, type, and management of postoperative complications. Abutment size, age, sex, indication for surgery, implant device type, duration of follow-up, and patient comorbidities were evaluated as potential factors affecting outcomes. Results: Average duration of follow-up was 16 months (range 6–29 mo). Postoperative complications occurred in 38 (29.2%) patients. Twenty-four (18.4%) patients experienced minor complications requiring simple, local care; eight (6.1%) patients required in-office procedural intervention; and six (4.6%) patients required revision surgery in the operating room. Implant extrusion occurred in three (2.3%) patients. Eleven (8.5%) patients required placement of a longer abutment. Patients receiving the 6 mm abutment at initial surgery were significantly more likely to encounter a complication requiring in-office procedural intervention or revision surgery (p = 0.001). Conclusion: Minor complications after implantation of percutaneous osseointegrated bone conduction devices are common. The vast majority of these complications are due to localized skin reactions, most of which are readily addressed through local care. Patients receiving the 9 mm abutment during initial implantation are significantly less likely to require in-office procedural intervention or revision surgery postoperatively as compared with those receiving the shorter, 6 mm abutment. Address correspondence and reprint requests to Sean R. Wise, MD, CDR, MC, USN, 34800 Bob Wilson Drive STE 200, San Diego, CA 92134; E-mail: srwise@earthlink.net The views expressed in this manuscript are those of the authors and do not necessarily reflect the official policy or position of the Department of the Navy, Department of Defense, or the United States government. The authors disclose no conflicts of interest. Copyright © 2018 by Otology & Neurotology, Inc. Image copyright © 2010 Wolters Kluwer Health/Anatomical Chart Company

from #Audiology via xlomafota13 on Inoreader http://ift.tt/2oM8OMz
via IFTTT

Δεν υπάρχουν σχόλια:

Δημοσίευση σχολίου