Normative Study of Wideband Acoustic Immittance Measures in Newborn Infants

Objective
The purpose of this study was to describe normative aspects of wideband acoustic immittance (WAI) measures obtained from healthy White neonates.
Method
In this cross-sectional study, wideband absorbance (WBA), admittance magnitude, and admittance phase were measured under ambient pressure condition in 326 ears from 203 neonates (M age = 45.9 hr) who passed a battery of tests, including automated auditory brainstem response, high-frequency tympanometry, and distortion product otoacoustic emissions.
Results
Normative WBA data were in agreement with most previous studies. Normative data for both WBA and admittance magnitude revealed double-peaked patterns with the 1st peak at 1.25–2 kHz and the 2nd peak at 5–8 kHz, while normative admittance phase data showed 2 peaks at 0.8 and 4 kHz. There were no significant differences between ears or gender for the 3 WAI measures. Standard deviations for all 3 measures were highest at frequencies above 4 kHz.
Conclusions
The 3 WAI measures between 1 kHz and 4 kHz may provide the most stable response of the outer and middle ear. WAI measures at frequencies above 4 kHz were more variable. The normative data established in the present study may serve as a reference for evaluating outer and middle ear function in neonates.

from #Audiology via xlomafota13 on Inoreader http://ift.tt/2ojBPjm